In most cases, the Cap itself is not the cause of a charging problem. The charging cable is the most common cause, so we recommend starting there.
Revian Red Caps use specialized power-only cables that do not contain data wires. Some third-party cables contain data wires that can interfere with the Cap and cause charging problems or inaccurate status readings. For this reason, always use the original Revian charging cable with your Revian System.
Important: Third-party cables may cause charging or operational issues, even if they appear to fit correctly.
While the Cap is plugged in, check the blue light near the charging port:
Charging: The blue light remains steadily illuminated.
Fully charged: The blue light blinks slowly.
If the blue light does not appear while the Cap is plugged in, please contact Revian Support.
Check the charging cable for fraying, bent connectors, or other visible damage.
Plug the charger into a different wall outlet to rule out a problem with the original outlet.
Make sure you are using the original wall adapter that came with your Cap. It should be rated 120V AC to 5V DC, 1A USB. A lower-powered adapter may prevent the Cap from charging properly.
When the adapter is connected to power, confirm that the Revian AC adapter displays a solid green light.
If you have completed the steps above, select the type of charging-port connector used by your Cap for additional troubleshooting instructions.
Check the Charging Port Connection Inside the cap’s charging port, there is a small oval metal charging tip. Please make sure it is fully inserted and seated properly.